The Laboratory Muffle Furnace YR05284-1 is an indispensable tool for any research lab demanding precision, safety, and efficiency in high-temperature environments. Its robust 64-liter chamber and innovative cooling ventilation system ensure safe operation while optimizing energy consumption, offering up to 80% savings compared to conventional models. Equipped with high-purity alumina fiber insulation and a sophisticated 30-segment PID controller for precise temperature control, it ensures protection against overheating. Designed for advanced research and industrial applications, this furnace stands out as a reliable and durable choice.
Market Price
The YR05284-1 Muffle Furnace is categorized within the mid to high-end range, reflecting its superior quality, durability, and precision. This makes it a valuable investment for research laboratories and industrial applications where reliability and performance are non-negotiable. For specific pricing details that accommodate your unique needs, you can swiftly and conveniently request a personalized quote via the Kalstein Plus platform.
Frequently Asked Questions
- What is the maximum working temperature?
The maximum temperature the furnace can achieve is 1700°C, but it is recommended to maintain a continuous working temperature of around 1600°C for optimal longevity. - What kind of protection does it include against overheating?
The furnace is equipped with automatic overheating protection and thermocouple failure monitoring, enhancing safety in controlled environments. - How is temperature control managed?
It features a PID system with 30 programmable segments, ensuring stability and precision with a temperature accuracy of +/- 1°C, perfect for high-accuracy research.
Advantages and Disadvantages
The YR05284-1 offers a myriad of advantages, including its durable construction, significant energy savings, and precise temperature control. Its large chamber is perfect for high-volume work, and its MoSi2 heating system guarantees uniform heat distribution. However, its size and weight may pose limitations for smaller labs, and for temperatures above 1400°C, a controlled heating rate of 2°C/min is advised to ensure durability.
Field Applications
This muffle furnace is critical for applications in analytical chemistry, metallurgy, and materials science. It is ideally suited for activities such as the heat treatment of ceramics, resistance testing, and sintering—scenarios where stable and controlled temperatures are essential. Thanks to its advanced safety and precision features, it supports materials research, ensuring experimental success.
Recommendations for Optimal Use
To enhance the lifespan of the YR05284-1, it is advisable to maintain a heating rate of under 10°C/min for operations below 1400°C, with specific rates adhered to at higher temperatures. Regular maintenance, including chamber cleaning and thermocouple checks, is vital for ensuring ongoing precision and safety. Don’t hesitate to request a personalized quote today through the Kalstein Plus platform to discover how this professional-grade furnace can elevate your research initiatives.
Product Features
- Model: YR05284-1
- Rated Voltage: AC 380V 3 phase or 1 phase 50/60Hz
- Power: 18kw
- Chamber Size: 400x400x400mm, 64L
- Max. Temperature: 1700°C
- Continuous Working Temperature: 1600°C
- Max. Heating Rate: 30°C/min; recommended rates: 10°C/min under 1400°C, 5°C/min at 1400°C-1600°C, 2°C/min above 1600°C
- Temperature Controller: PID automatic control with 30 programmable segments
- Temperature Accuracy: +/- 1°C
- Heating Element: MoSi2 heater
- Thermocouple: B type
- Standard Parts: Reserve heater 2pcs, crucible tongs 1pc, asbestos gloves 1pair, one user manual
Technical Specifications
Model | YR05284-1 |
Rated voltage | AC 380V 3 phase or 1 phase 50/60Hz |
Power | 18kw |
Chamber size | 400x400x400mm, 64L |
Max. Temperature | 1700 °C |
Continuous working temperature | 1600 °C |
Max. Heating Rate | 30 °C /min; Suggest to use 10°C/min under 1400°C, 1400°C-1600°C 5°C/min, >1600°C 2°C/min; |
Temperature controller | PID automatic control and auto-tune function. 30 programmable segments for precise control. Built in protection for overheating and broken thermocouple |
Temperature Accuracy | +/- 1°C |
Heating Element | MoSi2 heater |
Thermocouple | B type |
Standard parts | Reserve heater 2pcs, crucible tongs 1pc, asbestos gloves 1pair , one user manual |