Description
The Infant Incubator YR02186 is an advanced medical device engineered to provide optimal care and monitoring for newborns in critical care settings. With features like air and skin temperature servo-controls, this incubator guarantees precise thermal regulation, essential for maintaining an ideal environment for neonates. It includes multiple self-check alarms for enhanced safety and dependability. Furthermore, its adjustable mattress and removable humidity trough increase the usability and convenience for healthcare professionals.
Market Price
The market price for high-quality infant incubators such as the YR02186 can vary, reflecting the advanced technology and robust capabilities of the device. Prospective buyers are invited to generate an automatic quote using the Kalstein Plus platform to obtain up-to-date pricing information.
Frequently Asked Questions
How does the Infant Incubator YR02186 ensure the safety of newborns?
It features self-check alarms and precise temperature controls to maintain a safe environment.
Can the humidity levels be adjusted?
Yes, it has a humidity display and control function for fine-tuning the environment as needed.
Is there an option for phototherapy?
Yes, an optional phototherapy unit can be included if necessary.
Advantages and Disadvantages
Advantages: The incubator offers high accuracy in temperature control and a low noise level of ≤55dB, fostering a calm environment conducive to neonatal care. Additional features like a guardrail and optional electric weighing function enhance its usability.
Disadvantages: The advanced features may result in a higher initial investment compared to simpler models, but these benefits often outweigh the costs in critical care settings.
Product Use in the Field
In practical scenarios, the Infant Incubator YR02186 is employed in neonatal intensive care units to ensure a stable environment for premature or at-risk newborns. Healthcare providers depend on its precise temperature management and comprehensive alarm systems to deliver superior care. Its design also facilitates easy cleaning and maintenance, making it a favored choice in busy medical environments.
Recommendations
To maximize the longevity and performance of the incubator, it is recommended to perform regular maintenance and periodic checks of all electronic systems. Leverage the Kalstein Plus platform for support resources, procuring replacement parts, and exploring optional accessories.
Features
- Air and skin temperature servo-controlled by computer
- Various and self-check alarms
- Free step mattress tilting adjustment
- Removable humidity trough, easy to clean
- Temperature setting above 37°C
- Equipped with transfusion shelf and tray
- Added with guardrail
- RS-232 connector
- Humidity display and control function
- Optional phototherapy unit
Technical Specifications
Model | YR02186 |
Electrical requirement | ~220V 59Hz or ~230V 60Hz |
Power consumption | ≤450VA |
Air mode temperature range | 25.0°С~37.0°C |
Skin mode temperature range | 25.0°C~37.0°C |
Air over temperature range | <38.0°C |
Air mode high temperature range | 37.1°C~39.0°C |
Skin mode high temperature range | 37.1°C~38.0°C |
High temperature mode alarm | < 40.0°C |
Air mode alarm | ±3.0°C |
Skin mode alarm | ±1.0°C |
Skin sensor accuracy | ≤±3.0°C |
Air control accuracy | ≤0.5°C |
Mattress temperature uniformity | ≤0.8°C |
Noise | ≤55dB (A) |